What plastic material are compatible with sterilisation by irradiation?
Polymer compatibility with ionizing radiation depends on whether molecular chain scission (degradation, embrittlement) or cross-linking (stiffening, loss of elongation) dominates, as well as the presence of radiation stabilizers (e.g., Hindered Amine Light Stabilizers / HALS, mobilizing additives).
Gamma Co60: Operates at a low dose rate~1 to 10 kGy/hr over several hours. Atmospheric oxygen diffuses deep into the polymer matrix during processing, accelerating oxidative degradation and long-term shelf-life embrittlement.
E-Beam: Operates at an ultra-high dose rate >10^3 kGy/min completed in seconds. Dissolved oxygen inside the matrix is consumed almost instantly, limiting oxidative scission and giving oxygen-sensitive polymers higher functional thresholds than under gamma.
Standard medical single-use sterilization targets 25 kGy Dmin, with maximum doses Dmax frequently reaching 40–50kGy depending on the Dose Uniformity Ratio (DUR).

Medical Plastics Radiation Compatibility Table
Polymer Family | Common Medical Applications | Dominant Mechanism | Max Usable Dose: Gamma (kGy) | Max Usable Dose: E-Beam (kGy) | Primary Failure Modes & Processing Notes |
Aromatic Styrenics (PS, SAN, ABS) | Pipettes, petri dishes, roller bottles, luer fittings, housings | Cross-linking (protective aromatic rings) | >1,000 | >1,000 | High intrinsic radiation resistance; ABS high-impact grades may show minor loss of impact strength above 100kGy. Minor yellowing at >50 kGy. |
Aromatic Polyesters (PET, PETG, PBT) | Blister trays, balloon catheters, suture packages | Cross-linking | 500 –1,000 | >1,000$ | Retains clarity, tensile strength, and impact resistance. Moisture causes hydrolytic degradation during processing; dry thoroughly before molding. |
High-Performance Aromatics (PEEK, PSU, PES, PEI) | Surgical instruments, implants, dialyzer housings | Cross-linking / Resonance energy dissipation | >1,000 | >1,000 | Withstands repeated multi-cycle sterilization >100 kGy total dose). Polysulfones exhibit natural amber shift that deepens slightly. |
Polyurethane (TPU - Ether & Ester) | Vascular access catheters, tubing, wound dressings | Cross-linking | 100–1,000 | >1,000 | Aromatic TPUs resist radiation well; aliphatic grades show slight yellowing. Ether-based TPUs resist hydrolytic cleavage post-irradiation better than ester grades. |
Polyethylene (LDPE, LLDPE, HDPE) | Syringe plungers, packaging films, containers | Cross-linking | 100 –500 | 250 –1,000 | Gains tensile strength due to cross-linking, but loses percent elongation. Lower-density grades retain flexibility better than high-density variants. |
UHMWPE (Implant Grade) | Orthopaedic bearings (acetabular cups, tibial inserts) | Cross-linking (controlled) | 50 –100 | 50 –100 | Cross-linking improves wear resistance, but trapped free radicals cause fatigue cracking. Requires post-irradiation remelting/annealing or vitamin E stabilization. |
Polycarbonate (PC) | Dialyzers, stopcocks, oxygenators, centrifuge bowls | Chain scission & Cross-linking | 50 –100 | 100 –250 | Susceptible to noticeable yellow/amber discoloration at standard doses (25–40kGy). Retains mechanical strength, but requires tinted formulations to mask optical shift. |
Plasticized PVC (DEHP/TOTM) | IV bags, blood bags, peristaltic tubing | Cross-linking / Dehydrochlorination | 50 –100 | 100–200 | Releases trace amounts of hydrogen chloride (HCl) gas; causes yellow/brown tinting. Stabilizer and plasticizer packages are required to prevent tackiness and odour. |
Polyamides (Nylon 6, 6/6, 11, 12) | Catheter shafts, braided tubing, non-absorbable sutures | Scission & Cross-linking | 30–50 | 50 –100 | Loses elongation, exhibits surface chalking and discoloration at doses >50 kGy. Thin wall sections and sutures embrittle rapidly under gamma. |
Radiation-Stabilized Polypropylene (PP) | Disposable syringes, centrifuge tubes, pipette tips | Scission (moderated by HALS additives) | 30 –50 | 50 –80 | Requires mobilizing oils and hindered amine stabilizers. Immediate functional retention is often acceptable, but vulnerable to post-irradiation aging over 1–3 years. |
Natural / Unstabilized PP (Homopolymer) | Standard molded parts, commodity caps/connectors | Rapid Chain Scission | <20 | <25 | Not recommended. Severe embrittlement, catastrophic cracking during handling, and sharp drop in impact strength upon storage. |
Elastomers: Silicones (LSR, RTV) | O-rings, seals, septums, silicone tubing | Cross-linking | 50 –100 | 100 –200 | Increases durometer (hardness) and reduces elasticity. Platinum-cured silicones maintain elasticity significantly better than peroxide-cured systems. |
Elastomers: Fluoroelastomers (FKM) | Gaskets, dynamic seals, fluid handling | Chain scission | 25 –50 | 50 –75 | Softens, swells, and exhibits permanent compression set under high doses. Multiple sterilization cycles should be avoided. |
Cellulosics (Cellulose Acetate, Cellophane) | Dialysis membranes, paper packaging, filters | Rapid Chain Scission | 20 –40 | 40 –50 | Loss of bursting strength, embrittlement, and noticeable yellowing. Dialysis membrane flux rates degrade significant. |
Polyacetal / POM (Delrin®, Celcon®) | Precision gears, insulin pen mechanisms, stapler components | Severe Chain Scission | <5 –10 | <15 | Contraindicated. Catastrophic degradation at sub-sterilization doses; releases formaldehyde gas and crumbles under mechanical loads. |
Fluoropolymers (PTFE / Teflon®) | Catheter liners, vascular grafts, introducer sheaths | Catastrophic Chain Scission | <5 | <10 | Contraindicated. Tertiary carbon-fluorine bonds undergo immediate scission in the presence of trace oxygen, causing total loss of tensile and yield strength. |
Critical Engineering Takeaways about sterilisation y irradiation
The Dose-Rate Effect on Polymer Life: When testing polymers near their threshold limits (such as polyamides, plasticized PVC, or stabilized PP), E-beam provides wider margins of safety than gamma. The brief exposure time in E-beam suppresses the propagation of oxygen-dependent free-radical cascades.
Shelf-Life Validation (ASTM F1980): Post-irradiation radical reactions in semi-crystalline materials (PP, PE) mean that physical testing immediately after exposure is insufficient. Accelerated aging protocols must be accompanied by real-time shelf-life testing to ensure packaging seal integrity and mechanical performance do not fail over time.
Resterilization Risk: Highly radiation-resistant thermoplastics (ABS, PEEK, PET) can withstand secondary cycles. However, aliphatic polyamides, elastomers, and polypropylenes accumulate irreversible structural damage, making them unsuitable for multiple cycles.