What plastic material are compatible with sterilisation by irradiation?

Writer: Patricia Vest
Patricia Vest
Oct 19
4 min read

Polymer compatibility with ionizing radiation depends on whether molecular chain scission (degradation, embrittlement) or cross-linking (stiffening, loss of elongation) dominates, as well as the presence of radiation stabilizers (e.g., Hindered Amine Light Stabilizers / HALS, mobilizing additives).   

  • Gamma Co60: Operates at a low dose rate~1 to 10 kGy/hr over several hours. Atmospheric oxygen diffuses deep into the polymer matrix during processing, accelerating oxidative degradation and long-term shelf-life embrittlement.

  • E-Beam: Operates at an ultra-high dose rate >10^3 kGy/min completed in seconds. Dissolved oxygen inside the matrix is consumed almost instantly, limiting oxidative scission and giving oxygen-sensitive polymers higher functional thresholds than under gamma.   


Standard medical single-use sterilization targets 25 kGy Dmin, with maximum doses Dmax frequently reaching 40–50kGy depending on the Dose Uniformity Ratio (DUR).   

Line of 4 bottles on a table emitting vapours
Plastic bottles emitting vapours

Medical Plastics Radiation Compatibility Table

Polymer Family

Common Medical Applications

Dominant Mechanism

Max Usable Dose: Gamma (kGy)

Max Usable Dose: E-Beam (kGy)

Primary Failure Modes & Processing Notes

Aromatic Styrenics (PS, SAN, ABS)

Pipettes, petri dishes, roller bottles, luer fittings, housings

Cross-linking (protective aromatic rings)

>1,000

>1,000

High intrinsic radiation resistance; ABS high-impact grades may show minor loss of impact strength above 100kGy. Minor yellowing at >50 kGy.

Aromatic Polyesters (PET, PETG, PBT)

Blister trays, balloon catheters, suture packages

Cross-linking

500 –1,000

>1,000$

Retains clarity, tensile strength, and impact resistance. Moisture causes hydrolytic degradation during processing; dry thoroughly before molding.

High-Performance Aromatics (PEEK, PSU, PES, PEI)

Surgical instruments, implants, dialyzer housings

Cross-linking / Resonance energy dissipation

>1,000

>1,000

Withstands repeated multi-cycle sterilization >100 kGy total dose). Polysulfones exhibit natural amber shift that deepens slightly.

Polyurethane (TPU - Ether & Ester)

Vascular access catheters, tubing, wound dressings

Cross-linking

100–1,000

>1,000

Aromatic TPUs resist radiation well; aliphatic grades show slight yellowing. Ether-based TPUs resist hydrolytic cleavage post-irradiation better than ester grades.

Polyethylene (LDPE, LLDPE, HDPE)

Syringe plungers, packaging films, containers

Cross-linking

100 –500

250 –1,000

Gains tensile strength due to cross-linking, but loses percent elongation. Lower-density grades retain flexibility better than high-density variants.

UHMWPE (Implant Grade)

Orthopaedic bearings (acetabular cups, tibial inserts)

Cross-linking (controlled)

50 –100

50 –100

Cross-linking improves wear resistance, but trapped free radicals cause fatigue cracking. Requires post-irradiation remelting/annealing or vitamin E stabilization.

Polycarbonate (PC)

Dialyzers, stopcocks, oxygenators, centrifuge bowls

Chain scission & Cross-linking

50 –100

100 –250

Susceptible to noticeable yellow/amber discoloration at standard doses (25–40kGy). Retains mechanical strength, but requires tinted formulations to mask optical shift.

Plasticized PVC (DEHP/TOTM)

IV bags, blood bags, peristaltic tubing

Cross-linking / Dehydrochlorination

50 –100

100–200

Releases trace amounts of hydrogen chloride (HCl) gas; causes yellow/brown tinting. Stabilizer and plasticizer packages are required to prevent tackiness and odour.

Polyamides (Nylon 6, 6/6, 11, 12)

Catheter shafts, braided tubing, non-absorbable sutures

Scission & Cross-linking

30–50

50 –100

Loses elongation, exhibits surface chalking and discoloration at doses >50 kGy. Thin wall sections and sutures embrittle rapidly under gamma.

Radiation-Stabilized Polypropylene (PP)

Disposable syringes, centrifuge tubes, pipette tips

Scission (moderated by HALS additives)

30 –50

50 –80

Requires mobilizing oils and hindered amine stabilizers. Immediate functional retention is often acceptable, but vulnerable to post-irradiation aging over 1–3 years.

Natural / Unstabilized PP (Homopolymer)

Standard molded parts, commodity caps/connectors

Rapid Chain Scission

<20

<25

Not recommended. Severe embrittlement, catastrophic cracking during handling, and sharp drop in impact strength upon storage.

Elastomers: Silicones (LSR, RTV)

O-rings, seals, septums, silicone tubing

Cross-linking

50 –100

100 –200

Increases durometer (hardness) and reduces elasticity. Platinum-cured silicones maintain elasticity significantly better than peroxide-cured systems.

Elastomers: Fluoroelastomers (FKM)

Gaskets, dynamic seals, fluid handling

Chain scission

25 –50

50 –75

Softens, swells, and exhibits permanent compression set under high doses. Multiple sterilization cycles should be avoided.

Cellulosics (Cellulose Acetate, Cellophane)

Dialysis membranes, paper packaging, filters

Rapid Chain Scission

20 –40

40 –50

Loss of bursting strength, embrittlement, and noticeable yellowing. Dialysis membrane flux rates degrade significant.

Polyacetal / POM (Delrin®, Celcon®)

Precision gears, insulin pen mechanisms, stapler components

Severe Chain Scission

<5 –10

<15

Contraindicated. Catastrophic degradation at sub-sterilization doses; releases formaldehyde gas and crumbles under mechanical loads.

Fluoropolymers (PTFE / Teflon®)

Catheter liners, vascular grafts, introducer sheaths

Catastrophic Chain Scission

<5

<10

Contraindicated. Tertiary carbon-fluorine bonds undergo immediate scission in the presence of trace oxygen, causing total loss of tensile and yield strength.

Critical Engineering Takeaways about sterilisation y irradiation

  1. The Dose-Rate Effect on Polymer Life: When testing polymers near their threshold limits (such as polyamides, plasticized PVC, or stabilized PP), E-beam provides wider margins of safety than gamma. The brief exposure time in E-beam suppresses the propagation of oxygen-dependent free-radical cascades.

  2. Shelf-Life Validation (ASTM F1980): Post-irradiation radical reactions in semi-crystalline materials (PP, PE) mean that physical testing immediately after exposure is insufficient. Accelerated aging protocols must be accompanied by real-time shelf-life testing to ensure packaging seal integrity and mechanical performance do not fail over time.

  3. Resterilization Risk: Highly radiation-resistant thermoplastics (ABS, PEEK, PET) can withstand secondary cycles. However, aliphatic polyamides, elastomers, and polypropylenes accumulate irreversible structural damage, making them unsuitable for multiple cycles.